BCI 314
8 VERSATILE INPUTS CONDITIONER TERMINAL BLOCK
FEATURES  
  • 8 isolated differential channels:
    bridge; 1/2 bridge; 1/4 bridge
  • Bridge impedance: 120Ω min.
  • Sensitivity: 3 ranges
    ± 5V ; ± 2.5V ; ± 1.25V ; ± 0,625V FS
    ± 500mV ; ± 250mV ; ± 125mV ; ± 62,5mVFS
    ± 100mV ; ± 50mV ; ± 25mV ; ± 12,5mV FS
  • Bridge and sensor power supply:
    5V or 10V / 50mA ; 12V or 24V / 50mA
  • Current sources: 4mA to 12mA
  • Compliance: 22V (ICP)
  • Coupling inputs: AC / DC
  • Front panel adjustable offset:
    ± 70 %, ± 50 % ± 20 % or ± 5 % FS
  • High common voltage ± 500V max
    CMRR : > 120dB at 60Hz
  • 8th order progressive elliptic “low pass” filter (option)
  • Differential outputs ± 10V FS
  • 6U / 5Te or 6Te format 300 series form factor
  • 7-pin LEMO connectors
  • Direct interface with ADAS boards

Adas board

 
DESCRIPTION

Developed within the framework of industrial conditioner products, the BCI 314 terminal block offers
8 configurable channels

  • voltage, current
  • Strain gages, potentiometer
  • Accelerometer sensor , etc. ...

For each channel:

  • the bridge excitation is + 5.000V or + 10.000V and the power supply is dimensioned to be able to supply 50mA with a regulation factor > 1/1000
  • A current source 4mA to 12 mA / 22V and an AC/DC coupling permit the use of “integrated electronic” sensors
  • Measurement is performed by an instrumentation amplifier with very low drift whose gains are selected with G1 = 1, 10, 50 and G2 = 2, 4, 8, 16 from which a ± 5V FS to ± 6.25mV FS


Each channel is” isolated”, High Common Mode ± 500V, which prevents ground return problems.

The BCI 314 terminal block replaces and improves products BCI 310 and BCI 312.

Afaq
9 rue Georges Besse – BP 47 – 78330 FONTENAY LE FLEURY – FRANCE
Tél.:(33) 1 30 58 90 09 - http://www.adas.fr - mail@adas.fr
Afaq
 

 

BCI 314

SPECIFICATIONS (t = 25°C)

TYPE

8 POLYVALENT INPUTS CONDITIONING TERMINAL BLOCK

INPUTS

 

- Number of channels

8 differential

- Level

± 5V FS (G1 = 1 ; G2 = 2) ; ± 10V with attenuation

- Impedance

ZIND =10MΩ

- Common mode rejection (CMRR)

> 86dB ( F ≤ 100Hz)

- Protection

± 50V max.

- Possible assembly modes

Bridge ; 1/2 bridge ; 1/4 bridge

- Value of "R"

Resistances on studs (Rmin = 120Ω )

- Gains (G = G1 x G2)

G1 selection: 1 ; 10 ; 50

 

G2 selection: 2, 4, 8, 16

- Input levels
(for output = ± 10V FS

G1

G2

2

4

8

16

1

± 5V

± 2,5V

± 1,25V

625mV

10

± 500mV

± 250mV

± 125mV

± 62,5mV

50

± 100mV

± 50mV

± 25mV

± 12,5mV

- Gains error G1

G1 = 10 → ± 0, 02 % FS

 

G1 = 50 → ± 0, 04 % FS

- Gains error G2

G2 = 2 to 16 → ± 0, 03 % FS

- Offset drift on input

≤ 2µV / °C

- Gain drift

≤ 50ppm / °C

- Linearity

≥ ± 0.02 % of FS " G

- Adjustable offset (front panel)

Coarse = ± 70 % ; ± 50 % ; 20 % of FS

(selection by strap)

Fine = 5 % of FS

- Adjustable gain (front panel) ± 5 % of FS

SUPPLIES SENSORS

 

BRIDGE EXCITATION

 

- Selectable level

+ 5.000V or + 10.000V

- Regulation

≥ 0.1 % 4-wire type

- Maximum current

50mA

- Stability

25 ppm/°C

ICP EXCITATION

 

- Current source

4 to 12mA adjustable

- Compliance

22V

POWER SUPPLY SENSOR

12V or 24V / 50 mA for voltage supply conditioner

ISOLATION

 

- Between channels

≥ 500VAC max

- Input/output (CMV)

± 500VAC max common mode

- IMR > 120dB @ 60Hz

OUTPUTS

 

- Type

Differential

- Level

± 10V FS

- Impedance

< 50 Ω Iout = 5mA

 

BCI 314

FILTERING

 

- High Pass Filter
- Low Pass Filter option

Coupling AC/DC inputs 0,1Hz 1st order HPF

SF

= Without Filter BW ≥ 25KHz (-3dB)

ALPF

= Adjustable Low Pass Filter

- Type
- Order
- Fc range
- Fc (-3dB)
- Control
- Fc accuracy
- Settling
- Bandpass ripple

- Attenuation
- THD + Noise

Progressive Elliptic
8th switching capacitors
100Hz to 12KHz (-0,7dB)
Fc x 1,04
FCLK = 100Fc
± 2 % of Fc
potentiometer
0dB to + 0,05dB - Dc to 0,5Fc
0dB to + 0,1dB - Dc to 0,8Fc
≥ 76dB at 2,5Fc
≤ 72dB

FLPF

= Fixed Low Pass Filter
- Type
- Fc range
- Fc accuracy
Butterworth 6th order
F3 = 100Hz ; F4 = 500Hz ; F5 = 1KHz ; F6 = 5KHz ; F7 = 10KHz ; F8 = 20 KHz
± 10%

CONNECTION

 

- Process
- System (plug-in / withdrawable)

7-pin LEMO connector for plug ref. : FGG1B307CLAD62
DIN 41612 64-pin female for BCI series backplane

POWER SUPPLY

 

- Voltage / current
- Protections

+ 24V rated / 1A (range 20V to 28V)
Delayed-action fuse and polarity reversal

PRESENTATION

 

- Format
- Dimensions in mm (L x W x H)
- Fitting

Double EUROPE / 5Te or 6Te
233.35 x 160 x 25,40
In 300 series rack (BCI RACK) or 9U enclosure

ENVIRONMENT

 

- Operating temperature
- Storage temperature
- Humidity

- 20°C to + 70°C
- 25°C to + 85 °C
90 % (without condensation)

EUROPEAN NORMS  
  EMC - EN 61326 - EN 55011 Class A
CE Compliance
ROHS - 2002/95/EC
HOW TO ORDER?

BCI 314- X / **** / ++

X
FRONT PANEL

****
Low Pass Filter

SF

ALPF

FLPF + ++
Without filter
8th Adjustable Low Pass Filter from 100Hz to 10KHz
6th Fixed Low Pass Filter

5 = 5Te

6 = 6Te

 
F3 = 100Hz
     
F4 = 500Hz
     
F5 = 1KHz
     
F6 = 5KHz
      F7 = 10KHz

   
F8 = 20KHz
ACCESSORIES

19”/6U enclosure to receive 12 or 16 BCIs series 300

BCI300-RACK

   
6
26/36
 
 
DOCUMENTATION Revision B - English Version
  Revision A - French Version
 
 
French Version